EasyManua.ls Logo

RADWAG MAX 60

RADWAG MAX 60
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 85 -
19. MEANS OF USING A MOISTURE ANALYZER FOR DRYING IN
TEMPERATURE OVER 160
°
C
In case of drying a sample in a temperature ranging between 161°C 250°C the
maximum time of maintaining the temperature is determined proportionally from the
range: 1 hour for 161 °C 20 minutes for 250°C.
When drying a sample in temperature 250°C, the maximum temperature value is
maintained by the instrument for 20 minutes, and then the software decreases the
temperature value to 160 °C without stopping the drying process.
Time interval of decreasing the drying temperature from 250 °C to 160 °C is 20
minutes.
In case of drying mode „Quick“ the maximum drying temperature is exceeded by 30
% but its value is not higher than the temperature value set in instrument’s factory
menu.
In case of drying mode „Step“ the maximum time interval of maintaining a set
temperature value in each step is limited to 20 minutes.
User manual no.
LMI-49-05/04/13/A

Table of Contents

Related product manuals