command on determining mass of an
empty pan, followed by a command to
place a weight with pre-determined
mass.
accept the process by pressing Enter
key. After completing the user
adjustment process the moisture
analyzer returns to submenu P1 -
As specified in the previous point return to the weighing mode.
It is recommended that the mass of an external adjustment weight is
approximately ¾ of the maximum capacity of a balance / moisture
analyzer.
8.1.3. Drying chamber (Temperature) adjustment
Temperature adjustment of the drying chamber is carried out using a dedicated
set available as additional equipment of the moisture analyzer.
1. Thermometer
2. Thermometer holder with a shield
Adjustment is a simple process aimed at adjusting the temperature sensor
assembled in moisture analyzer’s drying chamber. Before carrying out the