- 82 -
17.5. Error messages
Detecting of an error is signaled by the moisture analyzer by displaying a message
corresponding to the error. The message contains number of the error and a short
description informing on the possible causes of error occurrence.
List of error messages in a moisture analyzer:
Error message
Error
no.
Error description
"Control sum error"
1.1
Error related to data transmission.
"A/D error"
1.2
Converter error.
AD TcL error 1.3
The value of A/D divisions of actuator
thermometer = 0 temperature sensor
defected, return to weighing mode by
pressing ESC key.
AD TcH error 1.3
The value of A/D divisions of actuator
thermometer = 1024 temperature sensor
defected, return to weighing mode by
pressing ESC key.
AD ThL error 1.3
The value of A/D divisions of the
thermometer in the weighing chamber
= 0 temperature sensor defected heating
of weighing chamber shall not start,
return to weighing mode by pressing
ESC key.
AD ThH error 1.3
The value of A/D divisions of the
thermometer in the weighing chamber
= 1024 temperature sensor defected
heating of weighing chamber shall not
start, return to weighing mode by
pressing ESC key.
"Range exceeded " 2.1
Maximum measuring range of a balance
exceeded.
"Range exceeded " 2.2
Maximum measuring range of a balance
exceeded.
"A/D Null"
2.3
A/D converter error.
"A/D Full" 2.4
Maximum number of divisions from
converter exceeded.
"Tare/Zero out of range " 2.5
Permissible value of zeroing or tarring
for a balance is exceeded.
"Tare out of range " 2.6
Permissible value of tarring for a balance
is exceeded.