- 35 -
8.6.2. Non-standard printout templates
Principles of creating non-standard printout templates:
1. Moisture analyzer MAX series enables creating up to 4 custom
printout templates;
2. each template must have the string of text start and end specified,
e.g. Printout no. 1 Start – 1 and Printout no. 1 Stop 40. In this
case the Printout no. 1 contains text strings from 1 to 40.
3. Next insert the texts into the specified text strings, i.e. 1 ÷ 40.
It is recommended to use an external computer keyboard
connected to moisture analyzer’s port. This means of text input is
much simpler and quicker.
4. Non-standard printout templates may overlap each other, i.e.:
Printout 1 Start – 1
Printout 1 Stop 40
Printout 2 Start 20
Printout 2 Stop 40
A non-standard printout can contain:
Variables dependent on user needs (mass, date, project no., temperature
of the drying chamber, etc.)
Texts inserted in moisture analyzer’s user menu
A custom non-standard printout template can contain up to 640 characters inserted
in 80 text strings, 8 character each. (starting with parameter String 1 to String 80).
A balance user can create up to 4 non-standard printout templates.
8.6.3. Inserting text into strings
Variables in printouts
%%
Printout of a single character “%”
%K
Type of last completed
adjustment process
%N
Current net mass in basic measuring
unit
%I
Deviation in last completed
adjustment process
%d
Current date
%1
Code 1
%t
Current time
%2
Code 2
%i
Balance factory no.
%3
Code 3
%r
Software no.
%4
Code 4
%P
Project no.
%5
Code 5
%U
User no.
%6
Code 6
%k
Date and time of last completed
adjustment process

Table of Contents