The difference between initial zero value
and the factory value is more than 10 %
from (controlled after zero initiation).
The difference between the start mass
(mass of the drying pan and cross
shaped pan holder) and its value saved
in the memory is greater than 4 % on
moisture analyzer start (moisture
analyzer started with a load on the
"Difference > 1% Max" 2.9
The difference between mass of a
currently measured adjustment weight
and mass saved in balance memory is
greater than (difference >1%).
" Sample mass < 20gm " 2.10
The value of sample mass is below 20
mg.
”Sample mass out of
range "
2.11
Sample mass is below the value
determined in the parameter “Sample –
mandatory” in the database of drying
Value of a reference mass in Percent
Setup mode in smaller than 1000
The adjustment coefficient of the drying
chamber differs from the factory
adjustment coefficient by more than
Parameter value exceeds permissible
value.
Impermissible parameter value
Errors relating to data transmission to a
computer or a printer connected to a
moisture analyzer.
"CTS transmission
suspended "
" XOFF transmission
suspended"
"Time value exceeded " 6.1
Timeout error for carrying out a given
process (e.g. zeroing).
CAUTION:
If the carried out activities related to dealing troubleshooting or maintenance are
other than expected, please immediately contact you locla RADWAG dealer or
service point.