- 83 -
"Result > 10% Max" 2.7
The difference between initial zero value
and the factory value is more than 10 %
from (controlled after zero initiation).
"Result > 4% Max" 2.8
The difference between the start mass
(mass of the drying pan and cross
shaped pan holder) and its value saved
in the memory is greater than 4 % on
moisture analyzer start (moisture
analyzer started with a load on the
drying pan).
"Difference > 1% Max" 2.9
The difference between mass of a
currently measured adjustment weight
and mass saved in balance memory is
greater than (difference >1%).
" Sample mass < 20gm " 2.10
The value of sample mass is below 20
mg.
Sample mass out of
range "
2.11
Sample mass is below the value
determined in the parameter “Sample
mandatoryin the database of drying
programs.
„Ref < 1000 Div” 2.12
Value of a reference mass in Percent
Setup mode in smaller than 1000
reading units
Difference > 20%” 2.13
The adjustment coefficient of the drying
chamber differs from the factory
adjustment coefficient by more than
20%
"Out of range " 3.1
Parameter value exceeds permissible
value.
"Incorrect value "
3.2
Impermissible parameter value
"Incorrect password "
3.4
"Overwriting error "
4.1
Errors relating to data transmission to a
computer or a printer connected to a
moisture analyzer.
" Parity error "
4.2
" Frameerror "
4.3
"CTS transmission
suspended "
4.4
" XOFF transmission
suspended"
4.5
"Incorrect date "
5.1
Error of date setting.
"Time value exceeded " 6.1
Timeout error for carrying out a given
process (e.g. zeroing).
CAUTION:
If the carried out activities related to dealing troubleshooting or maintenance are
other than expected, please immediately contact you locla RADWAG dealer or
service point.