After assembling the adjustment
set close drying chamber’s lid
and start the temperature
adjustment process.
CAUTION: remember to carry out all the processes with great care not to
damage the mass measuring mechanism of a moisture analyzer.
Enter menu group P1 Adjustment.
Set the marker next to a parameter 04
Temperature adjustment.
Press RIGHT ARROW navigating key. The
moisture analyzer previews current
temperature value. After 8 minutes, the
display indicates flashing temperature value.
Enter the temperatrure value as indicated on
the external (control) thermometer.
Press Print/Enter key to proceeed to the
following step of drying chamber adjustment
process. The moisture analyzer switches on
the halogen heating element and for the
following 8 minutes it heats the drying
chamber, and simultaneously indicates on
its display the current temperature value.
After the 8 minute time interval the moisture
analyzer’s display indicates the flashing
temperature value. Enter the temperatrure
value as indicated on the external (control)
thermometer.
Press Print/Enter key to proceeed to the