EasyManua.ls Logo

RayTek MARATHON FA - 3 Technical Data; Measurement Specifications; FA Models; FR Models

RayTek MARATHON FA
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TechnicalData
4 MarathonSeriesFA/FR
3TechnicalData
3.1MeasurementSpecifications
3.1.1FAModels
TemperatureRange
FA1A 475to900°C(887°Fto1652°F)
FA1B 800to1900°C(1472°Fto3452°F)
FA1C 1200to3000°C(2192°Fto5432°F)
FA1G 750to1675°C(1382°Fto3047°F)
FA2A 250to800°C(482°Fto1472°F)
FA2B 400to1700°C(752°Fto3092°F)
SpectralResponse
FA1 1.0μm(Sidetector)
FA2 1.6μm(InGaAsdetector)
SystemAccuracy
1
FA1/FA2 ±(0.3%T
meas+2°C),Tmeasin°C
FA1G ±3°C
Repeatability ±1°C
TemperatureResolution
CurrentOutput ±0.05°C
±0.01°CforFA1G
DisplayandRS485 ±1°C
ResponseTime 10msec(95%),selectableto10sec
TemperatureCoefficient ±0.03%fullscalechangeper1°Cchangeinambient
temperature
NoiseEquivalentTemp.(NET) 1°Cpeaktopeak,attargetemissivity
of1.00
Emissivity 0.10to1.00,in0.01increments
SignalProcessing PeakHold,valleyhold,averaging
Holdtime0300sec,in0.1secincrements
3.1.2FRModels
TemperatureRange
FR1A 5001100°C(930°Fto2010°F)
FR1B 7001500°C(1290°Fto2730°F)
FR1C 10002500°C(1830°Fto4530°F)
SpectralResponse 1.0μmnominal(Si/Sisandwichdetector)

1
atambienttemperature23°C±5°C(73°F±9°F)

Table of Contents