Maintenance
MarathonSeriesFA/FR 45
10Maintenance
Oursales representatives and customer servicearealwaysatyourdisposal for questions regarding
applicationassistance,calibration,repair,andsolutionstospecificproblems.Pleasecontactyourlocal
salesrepresentativeifyouneedassistance.Inmanycases,problemscanbesolvedoverthetelephone.
If you need to return equipment for servicing,
calibration, or repair, please contact our Service
Departmentbeforeshipping.Phonenumbersarelistedatthebeginningofthisdocument.
10.1TroubleshootingMinorProblems
Symptom Probable Cause Solution
No output No power to instrument Check the power supply
Erroneous temperature Faulty sensor cable Verify cable continuity
Erroneous temperature Field of view obstruction Remove the obstruction
Erroneous temperature Window lens Clean the lens
Erroneous temperature Wrong slope or emissivity Correct the setting
Temperature fluctuates Wrong signal processing
Correct Peak/Valley Hold or Average settings
Table8:Troubleshooting