Harness connection between Xenon lamp and lamp regulator loose broken
defective
Harness between CIS and IPU loose, broken, defective
Harness between CIS and PSU loose, broken, defective
One or more defective: CIS, IPU, BCU, PSU, Xenon lamp
SC 144 B
CIS communication error
A serial communication error occurred with the CIS due to a PCB malfunction or
loose connection.
If cycling the machine power off/on does not solve the problem:
Harness between CIS and IPU loose broken defective
Harness between CIS and PSU loose, broken, defective
CIS defective
IPU defective
BCU defective
SC161 B
IPU Error
Three attempts to configure the Cetus chip on the IPU board failed when the machine
was switched on or when machine recovered from the energy save mode.
Replace the IPU
SC180 B
Left scanner fan lock
The BCU detected a left scanner fan lock signal 5 sec. after the machine was
switched on.
If cycling the machine power off/on does not solve the problem:
Left scanner fan harness connection to the IOB loose, broken, defective
Left scanner fan defective
IOB defective
BCU defective
SC181 B
Right scanner fan lock