4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG Page 27 6 - CALCULATION PROCEDURES
6.4.2 Calculation procedure 2 (C/F/Rb)
Method at which the difference of sample value A
S
and reagent blank A
RB
is multiplied with a given factor F.
The reagent blank A
RB
is entered or measured once.
Calculation procedure ....................................................... CP 2
Characteristic ............................................................ C / F / Rb
Method .................................................... End Point with Factor
Calculation formula ....................................... C = F (A
S
- A
RB
)
Factor ............................................................... given / entering
Reagent blank ........................................ entering or measuring
PHOTOMETER
4040+
# 1001
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 21: HDL-C
PROGRAM: 2
FACTOR: 325
WAVELENGTH: 546nm
TEMPERATURE: 37C
DELAY: 5s
UNIT: mg/dl
- - - - - - - - - - - -
MEASURE ZERO
Rb[A]: 0.058
NO. ABS. RESULT
1 1.064 327
2 1.188 367
3 1.340 417
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Insert / measure reagent blank
Insert / measure sample
Insert / measure sample
Insert / measure sample