EasyManua.ls Logo

Riele 4040+ - Calculation Procedure 8 (C;S;Sbrb)

Default Icon
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG Page 33 6 - CALCULATION PROCEDURES
6.4.8 Calculation procedure 8 (C/S/SbRb)
Method at which the difference of reagent blank A
RB
and blank of reagent blank A
RBB
regarding the amount is
subtracted of the difference of sample value A
S
and sample blank A
SB
regarding the amount and this difference
is multiplied with a factor F which is determined by measuring of a standard solution with known concentration
C
ST
and under consideration of standard blank A
STB
and the difference of reagent blank A
RB
and blank of reagent
blank A
RBB
The sample blank A
SB
is measured before every test. The reagent blank A
RB
is entered or measured once.
Calculation procedure ....................................................... CP 8
Characteristic ........................................................ C / S / SbRb
Method ............................................... End Point with Standard
Calculation formula .............. C = F ( |A
S
- A
SB
| - |A
RB
- A
RBB
| )
Resulting factor ................... F = C
ST
/ ( |A
ST
-A
STB
| - |A
RB
-A
RBB
| )
Reagent blank ........................................ entering or measuring
PHOTOMETER
4040+
# 1001
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 28: Ca
PROGRAM: 8
STANDARD: 8.02
WAVELENGTH: 546nm
TEMPERATURE: 37C
DELAY: 3s
MAX.UNITS: 12
UNIT: mg/dl
- - - - - - - - - - - -
MEASURE ZERO
Rb[A]: 0.150
Rbb[A]: 0.046
DELTA Rb: 0.104
ST[A] 1: 1.485
ST[A] 2: 1.521
ST[A] 3: 1.495
ST[A]: 1.501
STb[A]: 0.479
DELTA ST: 1.022
FACTOR: 8.74
NO. ABS. RESULT
1 1.495 7.89
Sb[A]: 0.489
2 1.542 7.89
Sb[A]: 0.535
3 1.394 8.39
Sb[A]: 0.329
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Insert / measure blank of reagent blank
Insert / measure reagent blank
(Resulting blank)
Insert / measure standard blank
Insert / measure standard 1
Insert / measure standard 2 (optional)
Insert / measure standard 3 (optional)
(Averaged standard)
(standard blank)
(Averaged standard minus standard blank)
(Resulting factor)
Insert / measure sample blank
Insert / measure sample
Insert / measure sample blank
Insert / measure sample
Insert / measure sample blank
Insert / measure sample

Table of Contents