ROBERT RIELE GmbH & Co KG Page 28 6 - CALCULATION PROCEDURES
6.4.3 Calculation procedure 3 (C/F/Sb)
Method at which the difference of sample value A
S
and sample blank A
SB
regarding the amount is multiplied with
a given factor F. The sample blank A
SB
is measured before every test.
Calculation procedure ....................................................... CP 3
Characteristic ............................................................ C / F / Sb
Method .................................................... End Point with Factor
Calculation formula ....................................... C = F |A
S
- A
SB
|
Factor ............................................................... given / entering
PHOTOMETER
4040+
# 1001
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 23: BILIRUBIN
PROGRAM: 3
FACTOR: 12.80
WAVELENGTH: 546nm
TEMPERATURE: 37C
DELAY: 5s
MAX.UNITS: 8.0
UNIT: mg/dl
- - - - - - - - - - - -
MEASURE ZERO
NO. ABS. RESULT
1 1.000 4.21
Sb[A]: 0.671
2 1.215 4.25
Sb[A]: 0.884
3 1.033 4.23
Sb[A]: 0.702
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Insert / measure sample blank
Insert / measure sample
Insert / measure sample blank
Insert / measure sample
Insert / measure sample blank
Insert / measure sample