4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG Page 31 6 - CALCULATION PROCEDURES
6.4.6 Calculation procedure 6 (C/S/Rb)
Method at which the difference of sample value A
S
and reagent blank A
RB
is multiplied with a factor F which is
determined by measuring of a standard solution with known concentration C
ST
and under consideration of reagent
blank A
RB
.
The reagent blank A
RB
is entered or measured once.
Calculation procedure ....................................................... CP 6
Characteristic ............................................................ C / S / Rb
Method ............................................... End Point with Standard
Calculation formula ....................................... C = F (A
S
- A
RB
)
Resulting factor ........................................... F = C
ST
/ (A
ST
-A
RB
)
Reagent blank ........................................ entering or measuring
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Insert / measure reagent blank
Insert / measure standard 1
Insert / measure standard 2 (optional)
Insert / measure standard 3 (optional)
(Averaged standard)
(Resulting factor)
Insert / measure sample
Insert / measure sample
Insert / measure sample