4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG Page 68 11 - TECHNICAL DATA
11.4 SHORT SPECIFICATIONS
Single-beam filter photometer
Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm
and 623 nm; 3 positions for optional filter of choice
Normal standard cuvettes
(macro or semi-micro, disposable or special optical glass)
Measuring cell as well as dry incubation block with 8 positions temperature
controlled by electrical thermostat at 37 °C
Touchscreen for direct functions and alphanumerical inputs
Graphic display: White characters or symbols, blue background, lighted,
resolution 240 * 128 dots
English and French/German/Indonesian/Russian/Spanish
General operating software can be updated by PC
Reagent-open system with capacity for up to 231 programmable methods
Data import by touchscreen or PC
Up to 50 nonlinear calibration curves with max 20 sets of points can be
stored
Serial port for connection to an external printer and/or PC
Up to 1000 results can be saved in the memory automatically
Absorbance
End point with factor, standard or multiple standards,
with or without reagent blank and/or sample blank
Bichromatic end point
Kinetics with factor, standard or multiple standards, with or without blank
Fixed time with factor, standard or multiple standards,
with or without reagent blank
Turbidimetry with optional timer function
Single, double and triple determinations
Curve fitting for nonlinear standard curves
Free hemoglobin in combination with optional interference filters
Up to 50 methods can be controlled with two control serums,
Levey Jennings plot
Kinetic: variable from 3 – 19 deltas, time per delta 3 – 255 s
Fixed time: variable from 0 – 1800 s
Programmable from 0 – 1800 s
Range: 100 V
AC
up to 240 V
AC
at 50/60 Hz
Length 33 cm x width 34 cm x height 18 cm