ROBERT RIELE GmbH & Co KG Page 35 6 - CALCULATION PROCEDURES
6.4.10 Calculation procedure 10 (FTK/S/Rb)
Method at which a reagent blank is measured after an incubation time ( A
RB,0
) and after a reaction time (
A
RB,1
) and also a sample after an incubation time ( A
S,0
) and after a reaction time ( A
S,1
).
The difference from the change of the sample and the change of the reagent blank becomes multiplied with a
factor F which is determined by means of the change of standard solution |A
ST,0
-A
ST,1
| and the change of reagent
blank |A
RB,0
-A
RB,1
| during the reaction time and given concentration of standard. The reagent blank A
RB
is entered
or measured once.
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [ENTER].
After each measurement the next sample can be measured with [NEXT].
With [RESULT] it is possible to measure the same sample again.
Calculation procedure ..................................................... CP 10
Characteristic ........................................................ FTK / S / Rb
Method ............................................. Fixed Time with Standard
Calculation formula ......... C = F ( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
Resulting factor .............. F = C
ST
/ ( |A
ST,0
-A
ST,1
| - |A
RB,0
-A
RB,1
| )
Reagent blank ........................................ entering or measuring
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print-out of the method data
follows.
The measuring window is shown.
Method procedure:
Insert / measure zero solution
Without reagent blank (insert / measure optionally)
Insert / measure standard 1
Insert / measure standard 2 (optional)
Insert / measure standard 3 (optional)
(Averaged standard)
(Resulting factor)
Insert / measure sample
Insert / measure sample
Insert / measure sample