4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG Page 49 8- UTILITY PROGRAMS
8.2.2 Multi-standard functions
Before curve data of a method with multi-standard can be processed, the method must have been established in
the method editor (chapter 7 - METHOD EDITOR). A curve without corresponding method cannot be processed!
The term “curve no” has the same meaning as “method no”.
If using a multi-standard method later on, pay attention that all extinction values of the samples lie within the range
of the curve bases. Values outside of the extinction range cannot be calculated. In this case “+-is shown and
“<<< >>>” printed instead of the reading.
Fig. 7.2.2.1
Main window of multi-standard functions
[MEASURE CURVE] After query of a curve
number and the first standard the program
changes automatically to the method
selection window. There the given
parameters of the respective method can
be again controlled and/or changed. All
further standards are queried during the
following procedure.
The program for measuring the multi-
standards branches out automatically to the
calculation procedures as follows:
CP1 CP5
CP2 CP6
CP3 CP7
CP4 CP8
CP9 CP10
CP11 CP12
CP14 CP14
Print-out of curve:
For measuring in another standard
method than suggested, start measuring
multi-standards in each other selected
standard method (before measuring the first
standard) by [MODE] [MODE] [M-STD].
Curve number and standards are queried
accordingly.
[MEASURE CURVE] After inquiry of a
curve number start of measuring of standard
directly with a basic method which is based
on standard. This special mode ends
automatically after measuring or after
selection of a factor method. The standard
of the selected basic method represents the
first standard which can be measured. All
further standards are queried during the
measuring execution in the special mode.
[PRINT CURVE] After inquiry of the curve
number print-out on the internal printer or via
the serial interface.
For measuring in a multi-standard
method at least 2 bases with A and C must
be defined!
U T I L I T I E S S U B M E N U E
M E A S U R E C U R V E L I S T
P R I N T C U R V E
E D I T C U R V E
S H O W C U R V E E X I T