ROBERT RIELE GmbH & Co KG Page 65 10 - ERROR MESSAGE / CORRECTION
check sum error in the data record of the device basic
setting (status, ADC correction)
division by a too small value (< 0.001 A)
check filter; check standard;
check measuring solution
invalid calibration curve
setting to zero not possible (zero value is < 32768 cycle)
check LED; check filter;
check zero solution
setting to zero not possible (zero value is > 983039 cycle)
repeat optic adjustment (chapter 8.2.1);
check LED; check filter;
check zero solution
measure valid standard solution
no parameter memory vacant (too little memory for non-
linear methods)
delete a no longer actual nonlinear
method
select other method no.; delete a no
longer actual nonlinear method
check sum error in the parameter memory (nonlinear
method)
at calculation overflows in nonlinear method
check factor; check parameter
wrong input of deltas or time per delta
communication: wrong data format
communication: sent data not plausible / not interpretable
communication: mentioned module does not answer in a
certain time
check interconnecting cable;
check mentioned module
communication: overflow send/receive buffer
reduce amount of data at the
communication partner
remote control: wrong method number
external software problem
remote control: unknown command
external software problem
remote control: wrong data format
external software problem
check sum of operating system bank 0 damaged
check sum of operating system bank 1 damaged
timeout at reception from module 2
check sum error at reception from module 2
NAK at reception from module 2
filter position out of tolerance