RIGOL
DS6000 Users Guide
5-5
Trigger Coupling
Trigger coupling decides which kind of components will be transmitted to the trigger
circuit. Note to distinguish it with Channel Coupling.
DC: allow the DC and AC components to pass the trigger path.
AC: block all the DC components and attenuate signals lower than 8 Hz.
LF reject: block the DC components and reject the low frequency components
lower than 8 kHz.
HF reject: reject the high frequency components higher than 65 kHz.
Press MENU Set Up Coupling in the trigger control area (TRIGGER) at the
front panel to select the desired coupling type (the default is DC). Note that the
trigger coupling is invalid when the Trigger Type is Video, Pattern or Protocol.