Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800/E Series   
Min Search 
 
Search for and mark the peak with the minimum amplitude on the trace. 
 
 
Peak Peak 
 
Execute peak search and minimum search at the samt time and mark the results 
with delta pair markers. Wherein, the result of peak search is marked with the 
delta marker and the result of minimum search is marked with the reference 
marker. 
 
 
Cont Peak 
 
Enable or disable continuous peak search and the default is Off. When enabled, the 
system will always execute a peak search automatically after each sweep in order 
to track the signal under measurement.  
 
The difference between Cont Peak and Signal Track 
In Cont peak, the system always searches for the maximum in the current 
channel; while in Signal Track, the system will search for and mark the point 
(with no more than 3 dB variation in amplitude) near the marker before Signal 
Track is enabled, as well as set the frequency of this point as the center 
 
 
Search Para 
 
Define the conditions of peak search for various peak searches. A real peak should 
meet the requirements of both the “PK Excursn” and “PK Thresh”. 
 
1. PK Excursn  
Set the excursion between the peak and the minimum amplitude on both 
sides of it. Peaks whose excursions are beyond the specified excursion are 
treated as real peaks.