RIGOL Chapter 2 Functions of the Front Panel of GPSA
2-54 RSA3000E User Guide
Table 2-41 Span for OBW Measurement
Parameter
Remarks
Default
2 MHz
Range
10 Hz to 3 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
OBW span/100, Min = 1 Hz
Left/Right Arrow Key Step
Up/Down Arrow Key Step
at 1-1.5-2-3-5-7.5 step
6. Power Ratio
Sets the percentage the signal power takes up in the whole span power. You can
use the numeric keys, the knob, or arrow keys to modify this parameter; you can
also use the touch screen to modify the parameter.
Table 2-42 Power Ratio for OBW Measurement
Parameter
Remarks
Default
99%
Range
1% to 99.99%
Unit
%
Knob Step
0.01%
Left/Right Arrow Key Step
Up/Down Arrow Key Step
1%
For details about the Auto Couple and Meas Preset menus, refer to their
descriptions on Page
2-45.