RIGOL Chapter 2 Functions of the Front Panel of GPSA
2-22 RSA3000E User Guide
Trigger
Selects the trigger source and sets trigger-related parameters.
Trigger Source
Sets "Free Run", "External 1", "External 2", or "Video" to be the trigger source.
Free Run
The trigger conditions are met at any time, that is, the analyzer generates trigger
signals continuously.
External 1
In this mode, an external signal is input via the [TRIGGER IN] connector on the
rear panel. When the signal meets the set trigger conditions, trigger signals are
generated.
1. Slope
Sets the trigger polarity for External Trigger 1 to "POS" or "NEG".
2. Delay State
Enables or disables the trigger delay function. After the trigger delay function is
enabled, you can set the trigger delay time.
3. Delay Time
Sets the time interval during which the instrument waits to start the sweep
operation after the trigger signal that meets the trigger conditions is generated.
You can use the numeric keys, the knob, or arrow keys to modify this parameter;
you can also use the touch screen to modify the parameter.
Table 2-18 Trigger Delay Time
Parameter
Remarks
Default
1 μs
Range
0 μs to 500 ms
Unit
s, ms, μs, ns, ps
Knob Step
trigger delay/100, Min = 1 μs
Left/Right Arrow Key Step
Up/Down Arrow Key Step
at 1-1.5-2-3-5-7.5 step

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3000E Series and is the answer not in the manual?

Rigol RSA3000E Series Specifications

General IconGeneral
BrandRigol
ModelRSA3000E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals