Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000E User Guide 2-75
Mkr->Ref
Sets the reference level of the analyzer to the amplitude of the current marker.
If Position marker is selected, the reference level will be set to the amplitude of
the current marker.
Given a Delta marker, if the current marker is the reference marker, then the
reference level is set to the amplitude of the reference marker; if the current
marker is the Delta marker, then the reference level is set to the amplitude of the
Delta marker.
MkrΔ->CF
Sets the center frequency of the analyzer to the frequency difference between the
two Delta markers.
Only when "Delta" marker is selected, can this function be valid.
This function is invalid in zero span.
MkrΔ->Span
Sets the span of the analyzer to the frequency difference between the two Delta
markers.
Only when "Delta" marker is selected, can this function be valid.
This function is invalid in zero span.