EasyManuals Logo
Home>Rigol>Measuring Instruments>RSA3000E Series

Rigol RSA3000E Series User Manual

Rigol RSA3000E Series
380 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #241 background imageLoading...
Page #241 background image
Chapter 4 Functions of the Front Panel of EMI RIGOL
RSA3000E User Guide 4-7
Table 4-7 Att(Meter)
Parameter
Remarks
Default
0 dB
Range
0 dB to 50 dB
Unit
dB
Knob Step
preamp off, step = 1 dB
Left/Right Arrow Key Step
Up/Down Arrow Key Step
5 dB
RF Preamp(Meter)
Sets the meter preamplifier to be on or off. When the signal under test is a low-level
signal, turning on the preamplifier can reduce the displayed average noise level, so
that you can distinguish low-level signals from the noise. By default, the preamp gain
is 20 dB.
Y Axis Unit
Sets the unit of the Y-axis to dBm, dBmV, dBμV, Volts, or Watts. Wherein, dBm, dBmV,
and dBμV are for Log scale; Volts and Watts are for Linear scale. The default unit is
dBuV.
Remarks:
The conversion relations between units are as follows:










×=
0.001W
1
R
2
Volts
10lgdBm
(4-3)








=
V1
Volts
20lgVdB
µ
µ
(4-4)








=
1mV
Volts
20lgdBmV
(4-5)
R
2
Volts
Watts =
(4-6)
Wherein,
R
denotes the reference resistance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3000E Series and is the answer not in the manual?

Rigol RSA3000E Series Specifications

General IconGeneral
BrandRigol
ModelRSA3000E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals