Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000E User Guide 2-31
OffsetOp1Trace
result
−=
(2-13)
In the above formula, the parameter unit is dB.
5. Op1–Op2+Ref
Operand 1 (Op1) minus Operand 2 (Op2), and then plus the reference
(Reference). Then, saves the result to the destination trace. During the sweep,
the following formula is executed for each point:
ReferenceOp2Op1Trace
result
+= -
(2-14)
In the formula, the unit for Op1, Op2, and Reference is dBm.
6. Off
Disables the math function.
Note: The trace math functions are mutually exclusive. If you apply one math
operation to a certain trace, the last selected math operation function should be
disabled.
Op1
Selects Operand 1 to be used for the trace math functions. Operand 2 can be
configured with Trace1, Trace2, Trace3, Trace4, Trace5, or Trace6.
Op2
Selects Operand 2 to be used for the trace math functions. Operand 2 can be
configured with Trace1, Trace2, Trace3, Trace4, Trace5, or Trace6.
Note: The currently display trace does not involve in math operation.
Offset
Sets the log offset in the math operation function. The unit is dB.
Note: This menu is only valid when you select "Op1+Offset" or "Op1-Offset" to be
the math type.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3000E Series and is the answer not in the manual?

Rigol RSA3000E Series Specifications

General IconGeneral
BrandRigol
ModelRSA3000E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals