- 77 - GD-70D
SET 10. FLT PTRN
This is a setting screen of the
fault alarm activation. Do not
change the setting when the
detector is used in a normal
way, because it determines
how the detector functions.
(Auto-reset setting: "nL" by
default)
SET 10
FLT PTRN
MAINTENANCE
SET 11.AT FLOW
Set the flow rate auto-
adjustment.
Select either ON or OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
flow rate auto-adjustment is
activated.
SET 11
AT FLOW
MAINTENANCE
SET
ON
AT FLOW
MAINTENANCE
↓↑
OFF
AT FLOW
MAINTENANCE
SET 12. ZERO F
Set the zero follower when
the electrochemical type
(ESU), pyrolysis-particle type
(SSU), new ceramic type
(NCU), or NDIR type (IRU)
sensor is installed.
Select either ON or OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
zero follower function is
activated.
* When SSU is installed, you
must select ON.
SET 12
ZERO F
MAINTENANCE
SET
ON
ZERO F
MAINTENANCE
↓↑
OFF
ZERO F
MAINTENANCE
SET 13.ZERO 24F
This is a supplemental
setting screen for the above
zero follower function. (A
setting to determine whether
the first-24-hour zero follower
will be performed after the
power is turned on.)
Do not change the setting
when the detector is used in
a normal way.
(ON by default)
SET 13
ZERO 24F
MAINTENANCE
SET 14.S ASSIST
Set the sensitivity correction
when the electrochemical
type (ESU) is installed.
Select either ON or OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
sensitivity correction function
is activated.
SET 14
S ASSIST
MAINTENANCE
SET
ON
S ASSIST
MAINTENANCE
↓↑
OFF
S ASSIST
MAINTENANCE