CALCulate Subsystem FSE
1065.6016.12 6.62 E-16
CALCulate:X and CALCulate:UNIT Subsystem
The CALCulate:X and CALCulate:Unit subsystems define the units for vector signal analyzer mode and
power measurements.
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:X
:UNIT
:TIME
:UNIT
:ANGLe
:POWer
S | SYM
DEG | RAD
DBM | V | W | DB |
PCT | UNITLESS |
DBPW | WATT |
DBUV | DBMV | VOLT |
DBPT | DBUA | AMPere
DBUV_MHZ | DBMV_MHZ |
DBUA_MHZ | DBUV_M | DBUA_M |
DBUV_MMHZ | DBUA_MMHZ
Vector Signal Analysis
Vector Signal Analysis
:CALCulate<1|2>:X:UNIT:TIME S | SYM
This command selects seconds or symbols as an x-axis unit.
Example: ":CALC:X:UNIT:TIME S"
Features: *RST value: S
SCPI: device-specific
Mode: VA-D
:CALCulate<1|2>: UNIT:ANGLe DEG | RAD
This command selects the unit for angular measurement.
Example: ":CALC:UNIT:ANGL DEG"
Features: *RST- value: RAD
SCPI: device-specific
Mode: VA-D
:CALCulate<1|2>: UNIT:POWer DBM | V | W | DB | PCT | DBPT | UNITLESS | DBPW | WATT |
DBUV |DBMV | VOLT | DBUA | AMPere | DBUV_MHZ |DBMV_MHZ|
DBUA_MHZ | DBUV_M | DBUA_M |DBUV_MMHZ | DBUA_MMHZ
This command selects the unit for power.
Example: ":CALC:UNIT:POW DBM"
Features: *RST value: _
SCPI: device-specific
Modes: A, VA
The units DEG, RAD, S, and HZ are available only in the vector analyzer mode.
DBUV_MHZ and DBUA_MHZ denote the units DBUV/MHZ or DBUA/MHZ.