FSE CALibration Subsystem
1065.6016.12 6.63 E-16
CALibration Subsystem
The commands of the CALibration subsystem perform instrument calibrations.
COMMAND PARAMETERS UNIT COMMENT
CALibration
[:ALL]?
:BANDwidth
[:RESolution]?
:BWIDth
[:RESolution]?
:IQ?
:LDETector?
:LOSuppression?
:PPEak?
:SHORt?
:STATe
--
--
--
--
--
--
--
--
<Boolean>
--
--
--
--
--
--
--
--
--
query only
query only
query only
query only /
Vector Signal Analysis
query only
query only
query only
FSEM/FSEK only
query only
::CALibration[:ALL]?
This command performs a complete calibration of the instrument. A "0" is returned if the calibration
was successful.
Example: "CAL?"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
:CALibration:BANDwidth | BWIDth[:RESolution]?
This command performs a calibration of the filter bandwidths. A "0" is returned if the calibration was
successful.
Example: ":CAL:BAND?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALibration:IQ?
This command performs a calibration of the vector signal analyzer. A "0" is returned if the calibration
was successful.
Example: ":CAL:IQ?"
features: *RST value: -
SCPI: device-specific
Modes: VA, BTS, MS