CALibration Subsystem FSE
1065.6016.12 6.64 E-16
:CALibration:LDETector?
This command performs a calibration of the log module’s characteristic and of the detectors. A "0" is
returned if the calibration was successful.
Example: ":CAL:LDET?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALibration:LOSuppression?
This command performs a calibration of the local oscillator suppression. . A "0" is returned if the
calibration was successful.
Example: ":CAL:LOS?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is only valid by model ‘30 instruments or by instruments which have been retrofitted.
:CALibration:PPEak?
This command performs a calibration of the tracking YIG filter (preselector peaking). A "0" is
returned if the calibration was successful.
Example: ":CAL:PPE?"
Features: *RST value: -
SCPI: device-specific
Modes: A
This command is only valid by the models FSEM and FSEK.
:CALibration:SHORt?
This command performs a short calibration. A "0" is returned if the calibration was successful.
Example: ":CAL:SHOR?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALibration:STATe ON | OFF
This command determines whether (ON) or not (OFF) the current calibration data are taken into
consideration.
Example: ":CAL:STAT OFF"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS