FSE CONFigure-Subsystem
1065.6016.12 6.67 E-16
:CONFigure[:BTS]:LIMit:PPEak <numeric_value>
This command determines the phase error limits in degrees for the phase/frequency measurement
(peak value).
Example: ":CONF:LIM:PPE 66"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:LIMit:PRMS <numeric_value>
This command determines the phase error limits in degrees for the phase/frequency measurement
(mean value).
Example: ":CONF:LIM:PRMS 22"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: BTS
Betriebsart: BTS
CONFigure[:BTS]:LIMit:EVMRms<numeric_value>
This command determines the value in percent for the error limits of error vector magnitude
measurement with RMS weighting.
Example: ":CONF:LIM:EVMR 40"
Features: *RST value depending on the standard
SCPI: device-specific
Mode: BTS
CONFigure[:BTS]:LIMit:EVMPeak<numeric_value>
This command determines the value in percent for the error limits of error vector magnitude
measurement with PEAK weighting.
Example: ":CONF:LIM:EVMP 30"
Features: *RST value depending on the standard
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:LIMit:OSUPpress<numeric_value>
This command defines the limit for the origin offset suppression.
Example: ":CONF:LIM:OSUP 30"
Features: *RST value depending on the standard
SCPI: device-specific
Mode: BTS