CONFigure-Subsystem FSE
1065.6016.12 6.68 E-16
:CONFigure[:BTS]:LIMit:PERCentile<numeric_value>
This command determines the 95% percentile limits. The percentile defines the value which the EVM
may exceed 5% of all symbols at maximum.
Example: ":CONF:LIM:PERC 30"
Features: *RST value depending on the standard
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:LIMit:FREQuency <numeric_value>
This command determines the frequency error limits in ppm for the phase/frequency measurement.
Example: ":CONF:LIM:FREQ 36"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:LIMit:STANdard ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) limit values.
Example: ":CONF:LIM:STAN ON"
Feature: *RST value: ON
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:POWer:CLASs <numeric_value> | M1 | M2 | M3 | P1
This command defines the power class of the base station.
Parameter: <numeric_value> ::= 1 to 8 (P-GSM phase I/II, E-GSM, R-GSM, GSM850)
::= 1 to 4 (PCS1900, DCS1800 phase I/II/II+)
M1, M2, M3 ::= Power Classes for Micro BTS
P1 ::= Power Class for Pico BTS
Example: ":CONF:POW:CLAS 4"
Features: *RST value: 4 (P-GSM phase I/II, E-GSM, R-GSM, GSM850)
1 (DCS1800, PCS1900)
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:POWer:COUPled ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) level values.
Example: ":CONF:POW:COUP ON"
Feature: *RST value: ON
SCPI: device-specific
Mode: BTS