FSE CONFigure-Subsystem
1065.6016.12 6.69 E-16
:CONFigure[:BTS]:POWer:STATic 0 to 6
This command defines the static power control level of the base station.
Example: ":CONF:POW:STAT 3"
Features: *RST value: 0
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:POWer:DYNamic 0 to 15
This command defines the dynamic power control level of the base station.
Example: ":CONF:POW:DYN 5"
Features: *RST value: 0
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:POWer:EXPected <numeric_value>
This command enters directly the rated output level of the base station specified by the
manufacturer.
Example: ":CONF:POW:EXP 43DBM"
Features: *RST value: 46 dBm (P-GSM phase I/II, E-GSM, R-GSM)
43 dBm (DCS1800, PCS1900)
SCPI: device-specific
Mode: BTS
:CONFigure[:BTS]:POWer:LIMit <numeric_value>
This command defines the level for the selection of level-dependent limit lines.
Example: ":CONF:POW:LIM 65DBM"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: BTS
This command is only available for the setting :CONFigure[:BTS]:POWer:COUPled OFF.
:CONFigure[:BTS]:POWer:SINGle[:STATe] ON | OFF
This command switches single measurement of carrier power on and off.
Example: ":CONF:POW:SING ON"
Feature: *RST value: OFF
SCPI: device-specific
Mode: BTS