CONFigure-Subsystem FSE
1065.6016.12 6.72 E-16
:CONFigure[:BTS]:TXSupp ON | OFF
This command defines that an additional carrier suppression of min. 20dB is taken into account for
the measurement. If there is already suppression, a more sensitive setting of the instrument is
selected.
Example: ":CONF:TXS ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: BTS
For measurements in the RX-band the value is automatically set to ON.
:CONFigure[:BTS]:PRESet
This command resets the parameters for the standard selected to their default values (DEFAULT
SETTINGS).
Example: ":CONF:PRES"
Features: *RST value: --
SCPI: device-specific
Mode: BTS
This command is an event and has thus no query and no *RST value assigned.
:CONFigure[:BTS]:SWEeptime STANdard | AUTO
This command selects the sweep-time computing mode for the spurious measurement:
Example: ":CONF:SWE AUTO"
Feature: *RST value: STANdard
SCPI: device-specific
Mode: BTS
STANdard The computation of the sweep time is based on a worst-case estimation
AUTO The sweep time is reduced by a factor of 8 (assuming all slots are on).
CONFigure[:BTS]:MTYPe GMSK | EDGE
This command selects the modulation type (GMSK or EDGE (8PSK)).
Example: ":CONF:MTYP EDGE"
Features: **RST value: GMSK
SCPI: device-specific
Mode: BTS
CONFigure[:BTS]:STYPe NORMal | MICRo | PICO
This comman selects the type of base station.
Example: ":CONF:STYP PICO"
Features: *RST value: NORM
SCPI: device-specific
Mode: BTS