CONFigure-Subsystem FSE
1065.6016.12 6.78 E-16
:CONFigure[:MS]:MEASurement?
This command queries which measurement is currently set.
PFERror Phase-/Frequency Error
MACCuracy Modulation Accuracy
POWer Carrier Power
PTEMplate Power v. Time
MODulation Modulation Spectrum
SWITching Transient Spectrum
SPURious Spurious
Example: ":CONF:MEAS?" Answer: "PFER"
Features: *RST value: -
SCPI: device specific
Mode: MS
This command is a query and has therefore no *RST value assigned.
:CONFigure[:MS]:ARFCn <numeric_value>
This command selects the number of the transmission channel of the mobile.
Parameter: <numeric_value>::= 1 to 124 (P-GSM phase I/II)
0 to 124, 975 to 1023 (E-GSM)
0 to 124, 955 to 1023 (R-GSM)
512 to 885 (DCS1800 phase I/II/II+)
512 to 810 (PCS1900)
128 to 251 (GSM850)
Example: ":CONF:ARFC 67"
Features: *RST value: 1 (P-GSM phase I/II)
0 (E-GSM; R-GSM)
512 (DCS1800 phase I/II/II+)
512 (PCS1900)
128 (GSM850)
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:ARFCn:AUTO ONCE
This command selects automatically the transmission channel of the mobile.
Example: ":CONF:ARFC:AUTO ONCE"
Features: *RST value: -
SCPI: device-specific
Mode: MS