EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 137

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONFigure-Subsystem FSE
1065.6016.12 6.80 E-16
:CONFigure[:MS]:LIMit:PERCentile<numeric_value>
This command defines the limit for the 95% percentile.
Example: ":CONF:LIM:PERC 30"
Features: *RST value depending on the standard
SCPI: device-specific
Mode: BTS
:CONFigure[:MS]:LIMit:FREQuency <numeric_value>
This command determines the frequency error limits in ppm for the phase/frequency measurement.
Example: ":CONF:LIM:FREQ 36"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:LIMit:STANdard ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) limit values.
Example: ":CONF:LIM:STAN ON"
Feature: *RST value: ON
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:POWer:CLASs <numeric_value> | EG1 | EG2 | EG3
This command defines the power class of the mobile.
Parameter: <numeric_value> ::= 1 to 5 (P-GSM phase I)
2 to 5 (P-GSM phase II, R-GSM850)
2 to 5 (E-GSM; R-GSM)
1 to 2 (DCS1800 phase I)
1 to 3 (DCS1800 phase II/II+)
1 to 3 (PCS1900)
Example: ":CONF:POW:CLAS 4"
Features: *RST value: 2 (P-GSM phase I/II, E-GSM, R-GSM, R-GSM850)
1 (DCS1800, PCS1900)
SCPI: device-specific
Mode: MS

Table of Contents