FSE CONFigure-Subsystem
1065.6016.12 6.81 E-16
:CONFigure[:MS]:POWer:COUPled ON | OFF
This command switches between user-defined (OFF) and standard-defined (ON) level values.
Example: ":CONF:POW:COUP ON"
Feature: *RST value: ON
ON standard
OFF user-defined
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:POWer:LEVel 0 to 31
This command defines the power control level of the mobile.
Example: ":CONF:POW:LEV 5"
Features: *RST value: 2 (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
0 (DCS1800, PCS1900)
SCPI: device-specific
Mode: MS
:CONFigure[:MS]:POWer:LIMit <numeric_value>
This command defines the level for the selection of level-dependent limit lines.
Example: ":CONF:POW:LIM 65DBM"
Feature: *RST value: depending on standard
SCPI: device-specific
Mode: MS
This command is only available for the setting :CONFigure[:MS]:POWer:COUPled OFF.
:CONFigure[:MS]:POWer:EXPected <numeric_value>
This command enters directly the rated output level of the mobile.
Example: ":CONF:POW:EXP 43DBM"
Features: *RST value: GMSK:
39 dBm (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
30 dBm (DCS1800, PCS1900)
EDGE
33 dBm (P-GSM Phase I/II, E-GSM, R-GSM, GSM850)
30 dBm (DCS1800, PCS1900)
SCPI: device-specific
Mode: MS