CONFigure-Subsystem FSE
1065.6016.12 6.88 E-16
:CONFigure:SPURious:RANGe TXBand | OTXBand | RXBand | IDLeband | COMBined
This command selects the frequency range used for the measurement.
Example: ":CONF:SPUR:RANG OTX"
Features: *RST value: TXB
SCPI: device-specific
Modes: BTS, MS
TXBand TX-Band
OTXBand Not TX-Band
RXBand RX-Band (option FSE-K11 only)
IDLeband Idle band (option FSE-K10 only)
COMBined TX-Band ± 2 MHz (option FSE-K11 only)
:CONFigure:SPURious:STEP<1 to 26> ON | OFF
This command selects a subband of the selected band for a spurious measurement.
Each band is divided up into 1 to max. 26 subbands, which are selected by the numerical suffix
following STEP. A subband is selected for measurement by setting ON.
Example: ":CONF:SPUR:STEP24 ON"
Feature: *RST value: ON
SCPI: device-specific
Mode: BTS, MS
:CONFigure:SPURious:STEP:COUNt?
This command queries the number of subbands of the currently selected band for a spurious
measurement.
Example: ":CONF:SPUR:STEP:COUNT?"
Feature: *RST value: -
SCPI: device-specific
Mode: BTS, MS
This command is a query and has therefore no *RST value assigned.
:CONFigure:SPURious:ANTenna CONDucted | RADiated
This command selects the features of the measurement of spurious emmissions.
Example: ":CONF:SPUR:ANT RAD"
Feature: *RST value: COND
SCPI: device-specific
Mode:MS