EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 153

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DISPlay Subsystem FSE
1065.6016.12 6.96 E-16
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM[:FREQuency]:CENTer
<numeric_value>
This command shifts the zoomed frequency range to the specified center frequency.
Example: ":DISP:TRAC:X:ZOOM:CENT 1GHZ"
Features: *RST value: -- (depending on the current frequency setting)
SCPI: device-specific
Mode: A-F
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe] 10dB to 200dB
This command defines the display range of the Y-axis (level axis) with logarithmic scaling
(DISP:TRAC:Y:SPAC LOG).
Example: ":DISP:TRAC:Y 110dB"
Features: *RST value: 100dB
SCPI: device-specific
Mode: A
For linear scaling, (DISP:TRAC:Y:SPAC LIN | PERC) the display range is fixed and cannot be
set. The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:MODE ABSolute | RELative
This command defines the scale of the y-axis (absolute or relative).
Example: ":DISP:TRAC:Y:MODE REL"
Features: *RST value: ABS
SCPI: device-specific
Mode: A
As long as SYSTem:DISPlay is set to OFF, this command does not directly influence the screen.
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:Y[:SCALe]:RLEVel -200dBm to 200dBm
This command defines the reference level.
Example: ":DISP:TRAC:Y:RLEV -60dBm"
Features: *RST value: -20dBm
SCPI: conforming
Modes: A, VA
The numeric suffix in TRACe<1 to 4> is not significant.

Table of Contents