FSE DISPlay Subsystem
1065.6016.12 6.95 E-16
:DISPlay[:WINDow<1|2>]:TIME ON | OFF
This command switches on or off the screen display of date and time.
Example: ":DISP:TIME ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:RVALue <numeric_value>
This command defines the reference value for the X-axis of the measuring diagram.
Example: ":DISP:TRAC:X:RVAL 20SYM"
Features: *RST value: -
SCPI: device-specific
Mode: VA-D
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM ON | OFF
This command displays the zoomed current frequency range in the opposite window of the split
screen representation.
Example: ":DISP:TRAC:X:ZOOM ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A-F
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM[:FREQuency]:STARt
<numeric_value>
This command defines the start frequency of the zoomed display range. The value must lie between
start and stop frequency of the original range.
Example: ":DISP:TRAC:X:ZOOM:STAR 100MHZ"
Features: *RST value: -- (depending on the current frequency setting)
SCPI: device-specific
Mode: A-F
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM[:FREQuency]:STOP <numeric_value>
This command defines the stop frequency of the zoomed display range. The value must lie between
start and stop frequency of the original range.
Example: ":DISP:TRAC:X:ZOOM:STOP 200MHZ"
Features: *RST value: -- (depending on the current frequency setting)
SCPI: device-specific
Mode: A-F
The numeric suffix in TRACe<1 to 4> is not significant.