DISPlay Subsystem FSE
1065.6016.12 6.94 E-16
:DISPlay[:WINDow<1|2>]:SELect
This command selects the active measurement window via the numeric suffix in WINDow . This way,
a switch is possible from FULL SCREEN A to FULL SCREEN B (see example).
Example: ":DISP:FORM SPLit"
":DISP:WIND2:SEL"
":DISP:FORM SINGle"
Features: *RST value: --
SCPI: device-specific
Modes: A, VA
This command is an event and has thus no query and no *RST value assigned.
:DISPlay[:WINDow<1|2>]:MINFo ON | OFF
This command switches the marker info list on the screen on or off.
Example: ":DISP:MINF ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA, BTS, MS
:DISPlay[:WINDow<1|2>]:TEXT[:DATA] <string>
This command defines a comment (max. 50 characters) which can be displayed on the screen.
Example: ":DISP:TEXT "signal/noise power measurement"
Features: *RST value: "" (empty)
SCPI: conforming
Modes: A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.
:DISPlay[:WINDow<1|2>]:TEXT:STATe ON | OFF
This command switches on or off the screen display of the comment.
Example: ":DISP:TEXT:STAT ON"
Features: *RST value: OFF
SCPI: conforming
Modes: A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.