FSE FETCh-Subsystem
1065.6016.12 6.105 E-16
:FETCh:BURSt:POWer[:IMMediate]?
Carrier Power measurement (:CONFigure:MS:POWer:SINGle:STATe OFF):
This command reads out the result of the last step performed during the measurement of the output
power of the base station or mobile.
Parameter: The result is output as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-Level>, <Delta>,<Status> with
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>: current dynamic power control level
<Rat-Level>: Rated value for the current power control level acc. to
standard dBm
<Act-Level>: measured power in dBm
<Delta>: Difference between the measured power and the power at the previous
static/dynamic power control level.
<Status>: Result of limit check in character data form:
PASSED no limits exceeded
FAILED limit exceeded
Example: ":FETC:BURS:POW?"
Result: 0,0,43,44.1,0,PASSED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
Carrier Power Individual measurement (:CONFigure:MS:POWer:SINGle:STATe ON)
This command reads out the result of the last step performed during the measurement of the output
power of the base station or mobile.
Parameter: The result is output as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,Rat-Level>,<Act-Level>,<RBW>,<ARFCN>,<CF>,
<Attenuation>,<Number of bursts>,<Status>
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>: current dynamic power control level
<Rat-Level>: Rated value for the current power control level acc. to
standard dBm
<Act-Level>: measured power in dBm
<RBW>: resolution Bandwidth in kHz
<ARFCN>: channel number
<CF>: carrier frequency in in Hz
<Att>: external attenuation in dBm
<Number of burst>: number of bursts
<Status>: result of limit check in character data form:
PASSED no limits exceeded
FAILED limit exceeded
Example: ":READ:BURS:POW?"
Result: 0,3,37,20.6915,1000,2,8.904E+008,20,1,FAILED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if