FSE FETCh-Subsystem
1065.6016.12 6.107 E-16
:FETCh:BURSt:MACCuracy:RMS:STATus?
This command reads out the status of the RMS measurement of the modulation accuracy taken over
the selected number of bursts.
0: failed, 1: passed
Example: ":FETC:BURS:MACC:RMS:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:RMS:AVERage?
This command reads out the average of the RMS measurement of the modulation accuracy taken
over the selected number of bursts.
Example: ":FETC:BURS:MACC:RMS:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:RMS:MAXimum?
This command reads out the maximum of the RMS measurement of the modulation accuracy taken
over the selected number of bursts.
Example: ":FETC:BURS:MACC:RMS:MAX?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:PEAK:STATus?
This command reads out the status of the PEAK measurement of the modulation accuracy taken
over the selected number of bursts.
0: failed, 1: passed
Example: ":FETC:BURS:MACC:PEAK:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).