FETCh-Subsystem FSE
1065.6016.12 6.108 E-16
:FETCh:BURSt:MACCuracy:PEAK:AVERage?
This command reads out the average of the PEAK measurement of the modulation accuracy taken
over the selected number of bursts.
Example: ":FETC:BURS:MACC:PEAK:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:PEAK:MAXimum?
This command reads out the maximum of the PEAK measurement of the modulation accuracy taken
over the selected number of bursts.
Example: ":FETC:BURS:MACC:PEAK:MAX?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:OSUPpress:STATus?
This command reads out the status of the original offset supression measurement of the modulation
accuracy taken over the selected number of bursts.
0: failed, 1: passed
Example: ":FETC:BURS:MACC:OSUP:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:OSUPpress:AVERage?
This command reads out the average of the original offset supression measurement of the
modulation accuracy taken over the selected number of bursts.
Example: ":FETC:BURS:MACC:OSUP:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).