FETCh-Subsystem FSE
1065.6016.12 6.110 E-16
:FETCh:BURSt:MACCuracy:FREQuency:STATus?
This command reads out the status of the frequency error measurement of the modulation accuracy
taken over the selected number of bursts.
0: failed, 1: passed
Example: ":FETC:BURS:MACC:FREQ:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:FREQuency:AVERage?
This command reads out the average of the frequency error measurement of the modulation
accuracy taken over the selected number of bursts.
Example: ":FETC:BURS:MACC:FREQ:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).
:FETCh:BURSt:MACCuracy:FREQuency:MAXimum?
This command reads out the maximum of the frequency error measurement of the modulation
accuracy taken over the selected number of bursts.
Example: ":FETC:BURS:MACC:FREQ:MAX?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see CONFigure:BURSt:MACCuracy).