EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 168

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSE FETCh-Subsystem
1065.6016.12 6.111 E-16
FETCh:PTEMplate Subsystem
This subsystem provides the commands for reading out results of measurements in the GSM BTS
(FSE-K11) or GSM MS (FSE-K10) Analyzer mode which are used to determine the carrier power of ,
power versus time measurement without first restarting a new measurement.
Command Parameter Unit Comment
FETCh Option FSE-K11, FSE-K10
:PTEMplate
:REFerence? query only
:FETCh:PTEMplate:REFerence?
This command reads out the results of the premeasurement
Parameter: The result is output as a list of partial result strings separated by ,’ in the following
(ASCII) format:
<Level1>,<Level2>,<RBW>
<Level1>: measured level
<Level2>: level corrected by means of the bandwidth
<RBW>: bandwidth
Example: ":FETC:PTEM:REF?"
Result: 43.2,43.2,1000000
Features: *RST value: --
SCPI: devicespecific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of Power versus Time is selected (see :CONFigure:BURSt:PTEMplate).