FSE FETCh-Subsystem
1065.6016.12 6.113 E-16
Example: ":FETC:SPEC:MOD? TXB"
Result: 0,890E6,915E6,-87.4,-108.0,ABS,FAILED,
1,893.2E6,893.2E6,-83.2,-108.0,ABS,FAILED,
2,895.7E6,895.7E6,-87.4,-108.0,ABS,FAILED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
ARFCn ARFCN ± 1.8 MHz
TXBand TX-Band
RXBand RX-Band
COMBined ARFCN ± 1.8 MHz / TX-Band (option FSE-K11 only)
DCSRx1800 RX band DCS 1800 (option FSE-K10 only)
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation spectrum is selected (see
:CONFigure:SPECtrum:MODulation).
:FETCh:SPECtrum:MODulation:REFerence?
This command reads out the result of the premeasurement.
Parameter: The result is output as a list of partial result strings separated by ,’ in the following
(ASCII) format:
<Level1>,<Level2>,<RBW>
<Level1>: measured level
<Level2>: level corrected by means of the bandwidth
<RBW>: bandwidth
Example: ":FETC:SPEC:MOD:REF?"
Result: 36.2,43.2,30000
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation spectrum is selected (see
:CONFigure:SPECtrum:MODulation).