INSTrument Subsystem FSE
1065.6016.12 6.130 E-16
INSTrument Subsystem
The INSTrument subsystem selects the operating mode of the unit either via text parameters or fixed
numbers. In the split-screen representation, a distinction is made between INSTrument1 (screen A) and
INSTrument2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
INSTrument<1|2>
[:SELect]
:NSELect
:COUPled
SANalyzer | DDEMod | ADEMod |
BSGM | MGSM
<numeric_value>
NONE | MODE | X | Y | CONTrol |
XY | XCONtrol | YCONtrol | ALL
Vector Signal Analysis
FSE-K11 or FSE-K10,
:INSTrument<1|2>[:SELect] SANalyzer | DDEMod | ADEMod | BGSM | MGSM
This command switches between the operating modes by means of text parameters.
Parameter: SANalyzer: spectrum analysis
DDEMod: vector signal analysis, digital demodulation
ADEMod: vector signal analysis, analog demodulation
BGSM: GSM BTS analysis
MGSM: GSM MS analysis
Example: ":INST DDEM"
Features: *RST value: SANalyzer
SCPI: conforming
Modes: A, VA, BTS, MS
Switchover to DDEMod or ADEMod is only possible in conjunction with option FSE-B7, Vector Signal
Analysis.
Switchover to BGSM is only possible in conjunction with option FSE-K11, GSM BTS Analyzer
Switchover to MGSM is only possible in conjunction with option FSE-K10, GSM MS Analyzer
:INSTrument<1|2>:NSELect 1 to 5
This command switches between the two modes by means of numbers.
Example: ":INST:NSEL 2"
Features: *RST value: 1
SCPI: conforming
Modes: A, VA, BTS, MS
1: spectrum analysis
2: vector signal analysis, digital demodulation
3: vector signal analysis, analog demodulation
4: GSM BTS analysis
5: GSM MS analysis
Switchover to 2 or 3 is only possible in conjunction with option FSE-B7, Vector Signal Analysis.
Switchover to 4 is only possible in conjunction with option FSE-K11, GSM BTS Analyzer
Switchover to 5 is only possible in conjunction with option FSE-K10, GSM MS Analyzer