FSE INSTrument Subsystem
1065.6016.12 6.131 E-16
:INSTrument<1|2>:COUPle NONE | MODE | X | Y | CONTrol | XY | XCONtrol | YCONtrol | ALL
This command defines the coupling between the two measurement windows screen A and B.
Example: ":INST:COUP NONE"
Features: *RST value: ALL
SCPI: device specific
Modes: A, VA
NONE no coupling.
MODE the operating mode of the two screens is coupled.
X or Y the scaling of the x- or y axis of the two screens is coupled.
CONTrol the trigger and gate parameter ,and the sweep parameters SINGle/ CONTinous
and COUNt of the two screens are coupled.
XY the scaling of the x- and y-axis of the two screens are coupled.
XCONTrol bzw. the trigger and gate parameter, and the sweep parameters SINGle CONTinous/
YCONTrol COUNt of the two screens are coupled.
ALL the scaling of the x- or y axis the trigger and gate parameter and the sweep
parameters SINGle/ CONTinous/ COUNt of the two screens are coupled.
The numeric suffix in INSTrument<1|2> is not significant.