EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 194

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSE MMEMory Subsystem
1065.6016.12 6.137 E-16
:MMEMory:MSIS A: | C:
This command changes to the drive indicated.
Example: ":MMEM:MSIS A:’"
Features: *RST value: "C:
SCPI: conforming
Modes: A, VA, BTS, MS
The drive may be the internal hard disk C: or the floppy-disk drive A:. The drive is indicated according
to the DOS conventions.
:MMEMory:NAME <file_name>
This command specifies a file which is printed or plotted to.
Parameter: <file_name> ::= DOS filename
Example: ":MMEM:NAME PLOT1.HPG’"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
The file name includes indication of the path and may also include the drive. The file name and path
information correspond to the DOS conventions. The output to the printer is routed into a file using
the command ":HCOP:DEST MMEM".
:MMEMory:RDIRectory <directory_name>
This command deletes the directory indicated.
Parameter: <directory_name>::= DOS path name
Example: ":MMEM:RDIR C:\TEST’"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
The directory name includes indication of the path and may also include the drive. The path name
corresponds to the DOS conventions. This command is an event which is why it is not assigned an
*RST value and has no query.
:MMEMory:STORe:STATe 1,<file_name>
This command stores the current instrument setting in a file.
Parameter: <file_name> ::= DOS file name without extension
Example: ":MMEM:STOR:STAT 1,TEST’"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
The current instrument state is stored as a file. The file name includes indication of the path and may
also include the drive. The path name corresponds to the DOS conventions. This command is an
event which is why it is not assigned an *RST value and has no query.