MMEMory Subsystem FSE
1065.6016.12 6.138 E-16
:MMEMory:STORe:TRACe 1 to 4,<file_name>
This command stores the selected trace (1 to 4) in ASCII format in a file.
Parameter: 1 to 4 := selected trace 1 to 4
<file_name> := DOS file name
Example: ":MMEM:STOR:TRAC 3,A:\TEST.ASC’"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
The file name includes indication of the path and may also include the drive. The file name and path
information correspond to the DOS conventions.This command is an event which is why it is not
assigned an *RST value and has no query.
:MMEMory:CLEar:STATe 1,<file_name>
This command deletes the instrument setting denoted by <file_name>.
Parameter: <file_name> ::= DOS file name without extension
Example: ":MMEM:CLE:STAT 1,TEST’"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
The device data set specified is deleted. The file name includes indication of the path and may also
include the drive. The path name corresponds to the DOS conventions. This command is an event
which is why it is not assigned an *RST value and has no query.
:MMEMory:CLEar:ALL
This command deletes all instrument settings in the current directory.
Example: ":MMEM:CLE:ALL"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is an "event" which is why it is not assigned an *RST value and has no query.
:MMEMory:SELect[:ITEM]:GSETup ON | OFF
This command includes the data of the general setup in the list of data subsets of a device setting to
be stored/loaded.
Example: ":MMEM:SEL:GSET ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA, BTS, MS