FSE OUTPut Subsystem
1065.6016.12 6.143 E-16
OUTPut Subsystem
The OUTPut subsystem checks the output features of the instrument.
In conjunction with option tracking generator, in the split screen mode, a distinction is made between
OUTPut1 (screen A) and OUTPut2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
OUTPut<1|2>
[:STATe]
:UPORt<1|2>
[:VALue]
:STATe
:AF
:SENSitivity
<boolean>
<binary>
<boolean>
<numeric_value
--
--
--
PCT |
HZ | KHZ |
DEG | RAD
Option Tracking Generator
Vector Signal Analysis
:OUTPut<1|2>[:STATe] ON | OFF
This command switches the tracking generator on or off.
Example: ":OUTP ON"
Features: *RST value: -
SCPI: conforming
Modes:A, VA
This command is only valid in conjunction with one of the options tracking generator.
:OUTPut<1|2>UPORt<1|2>[:VALue] #B00000000 to #B11111111
This command sets the control lines of the user ports.
Example: ":OUTP:UPOR2 #B10100101"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
User port 1 or 2 is written with the given binary pattern. If the user port is programmed to INPut
instead of OUTPut, the output value is temporarily stored.
:OUTPut<1|2>UPORt<1|2>:STATe ON | OFF
This command switches the control line of the user ports between INPut and OUTPut.
Example: ":OUTP:UPOR:STAT ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA, BTS, MS
With ON, the user port is switched to OUTPut, with OFF to INPut.